Medical X-ray Tube XD3A

Type: Station anode x-ray tube Application: For general diagnostic x-ray unit Model: RT13A-2.6-100 equivalent to XD3A-3.5/100 Integrated high quality glass tube

Products Details

Description

This tube, RT13A-2.6-100 is designed for general diagnostic x-ray unit and available for a nominal tube voltage with self-rectified circuit. RT13A-2.6-100 tube has one focus. The integrated high quality tube with glass design has one super imposed focal spot and a reinforced anode. The high anode heat storage capacity ensures a wide range of applications for general diagnostic x-ray application. A special designed anode enables an elevated heat dissipation rate which leads to a higher patient throughput and a longer product life. A constant high dose yield during the entire tube life is ensured by the high density tungsten target. Ease of integration into system products is facilitated by extensive technical support.

Technical data

Nominal  Tube Voltage 105kV
Nominal Inverse Voltage 115kV
Nominal Focal Spot 2.6(IEC60336/1993)
Max Anode Heat Content 30000J
Target Angle 19°
Filament Characteristics 4.5A, 7.0±0.7V
Permanent Filtration Min. 0.8mmAl/50kv(IEC60522/1999)
Target Material Tungsten
Tube Current 50mA

Competitive Advantage

Elevated anode heat storage capacity and cooling Constant high dose yield Excellent lifetime

Applications

RT13A-2.6-100 is Portable Diagnostic stationary anode X-ray tube, designed for general diagnostic x-ray unit and available for a nominal tube voltage with self-rectified circuit.

RT13A-2.6-100

Cautions
Read the cautions before using the tube
X-ray tube will emit X –ray when it is energized with high voltage, Special knowledge
should be required and cautions need to be taken when handling it。
1. Only a qualified specialist with X-Ray tube knowledge should assemble, maintain
and remove the tube。
2. Sufficient care should be taken to avoid strong impact and vibration to the tube
because it is made of fragile glass。
3. Radiation protection of the tube unit must be sufficiently taken。
4. The minimum sorce-skin distance (SSD) and the minimum filtration should match the
regulation and meet the standard。
5. The system should have proper overload protection circuit , the tube may be
damaged due to only one overload operation。
6. When any abnormalities are found during operation , immediately switch off the
power supply and contact the service engineer。
7. if the tube is with a lead shield, to dispose of lead shield must meet government
regulations。