Bone Desimeter X-ray Tube Brand Bx-1

Type: Station anode x-ray tube Application: Especially designed for designated for bone densimeter x-ray system for radiography. Model: RT2-0.5-80 Equivalent to BRAND X-RAY BX-1 Integrated high quality glass tube

Products Details

Description

RT2-0.5-80 Stationary X-Ray Tube is especially designated for bone densimeter x-ray system and available for nominal tube voltage constant potential circuit RT2-0.5-80 tube has one focus. The integrated high quality tube with glass design has one super imposed focal spot and areinforced anode. The high anode heat storage capacity ensures a wide range of applications for bone densimeter x-ray application. A special designed anode enables an elevated heat dissipation rate which leads to a higher patient throughput and a longer product life. A constant high dose yield during the entire tube life is ensured by the high density tungsten target. Ease of integration into system products is facilitated by extensive technical support.

Competitive Advantage

Elevated anode heat storage capacity and cooling Constant high dose yield Excellent lifetime

Applications

RT2-0.5-80 Stationary X-Ray Tube is especially designated for bone densimeter x-ray system and available for nominal tube voltage constant potential circuit.

Detailed Images

KL32-0.5-80

  Maintaining seasoning schedule Prior to usage, season the tube in accordance with the seasoning schedule given below until the required tube voltage is attained. Example given – needs to revise by the manufacturer and specified in the data sheet of the part: Initial incoming seasoning and seasoning schedule for idle period (more than 6 months) Circuit: DC (Center grounded)   KL34老练图When the tube current is unstable in seasoning, immediately switch off the tube voltage and after an interval of 5 minutes or more, increase the tube voltage gradually from the low voltage while making sure that the tube current is stable. The withstand voltage performance of the tube unit will be lowered as the exposure time and the number of operation increase. Stain-like impact traces might appear on the x-ray tube target surface by slight discharge in the course of the seasoning. These phenomena are one process to recover the withstanding voltage performance at that time. Therefore, if it is in stable operation at the maximum tube voltage of seasoning subsequent to them, the tube unit can be used without any interference to its electrical performance which in use.  

TECHNICAL DATA

This tube is especially designed for Bone Mineral Densitometer x-ray system unit.
Nominal Tube Voltage(self-rectified)……..………………..….………..…….……………95kV
(constant potential)………………..….………..…….……………...................................80kV
Nominal Tube Current .……………………………..….……….…………………………........2mA
Nominal Operating Condition………………………….…….....................80kV@2mA@2.4V
Nominal Anode Cooling Rate……………………………………………………………….....160W
Anode Heat storage Capacity ………………………………………………………………......16kJ
Filament current………………………………………………………….….....1.95 A@2.4V±0.2V
Focal spot …………………………………………………………………......0.5 (IEC60336/2005)
Target Angle………………………………………………………………….…….…………..............12°
Effective X-ray Cone Beam……………………………………………………….…….......24°X90°
Target Material………………………………………………………………….……….........Tungsten
Cathode type……………………….…………………………………………........Tungsten filament
Dimensions ……………………………………………..….....125mm length by 35mm diameter
Weight…………………………………………………………………………............approx.145 grams

Detailed Charts

Contact us

Please feel free to give your inquiry in the form below We will reply you in 24 hours