X-Flex High Voltage HV Cable for X-ray Equipment SRX-Z75

Chat Now
Product Details

X-Flex High Voltage HV Cable for X-ray Equipment SRX-Z75


Description:

This 3-conductor with rubber insulated high voltage cable’s typical applications are as followings:

1、Medical x-ray equipment like standard x-ray, computer tomography and angiography equipment.

2、Industrial and scientific x-ray or electron beam equipment like electron microscopy and x-ray diffraction equipment.

3、Low power high voltage testing and measuring equipment.

68.jpg


Highlights:
1,Excellent high withstand voltage, highest quality standard
2,Best cable flexibility, easy to integrate
3,Removable PIN connectors, fast maintenance
4,Cable length can be customized for you


Technical Data

Number of conductor

3

Rated voltage

75kVDC

Routine test voltage (high voltage insulation)

120kVDC/10min

Routine test voltage (conductor insulation)

2kVACrms/1min

Maximum conductor current

1.5mm2:15A

Nominal outside diameter

17.0±0.5mm

Thickness of PVC jacket

1.0mm

Thickness of high voltage insulation

4.5mm

Diameter of core-assembly

4.5mm

Insulation resistance core to shield @20℃

≥1×1012Ω·m

Conductor insulation resistance@20℃

≥1×1012Ω·m

Max Conductor resistance bare cond.@20℃

10.5mΩ/m

Max Conductor resistance insul. cond. @20℃

12.2 mΩ/m

Max Shield resistance@20℃

15 .0mΩ/m

Max Capacitance between conductor and shield

165nF/km

Max Capacitance between ins. cond. and bare cord

344nF/km

Max Capacitance between insulated conductors

300nF/km

Cable Min bending radius(static insulation)

40mm

Cable Min bending radius(dynamic installation)

80mm

Operating temperature

-10℃~+70℃

Storage temperature

-40℃~+70℃

Net weight

351kg/km